Oxoid Antimicrobial Susceptibility Testing Disc Cefoxitin. An in vitro diagnostic medical device (IVD) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Thermo Fisher Scientific Australia & New Zealand 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2015-RN-01097-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2015-11-10
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    An internal technical investigation by oxoid limited, part of thermo fisher scientific, has confirmed that certain lots of oxoid antimicrobial susceptibility testing disc cefoxitin may contain discs that have insufficient concentration of antibiotic. it has been identified that the low concentration may impact on the performance of the disc against escherichia coli and neisseria gonorrhoeae quality control organisms resulting in lower than expected zones of inhibition. performance with the staphylococcus aureus qc organism is currently within specification.Continued use of this lot could result in quality control failures, delayed results or incorrect results reporting.
  • 조치
    End users are requested to inspect their stock for affected units. Affected units are to be removed from further use. The requirement for the review of patient results should be determined by the laboratory director. This action has been closed-out on 22/08/2016.

Device

  • 모델명 / 제조번호(시리얼번호)
    Oxoid Antimicrobial Susceptibility Testing Disc Cefoxitin. An in vitro diagnostic medical device (IVD) Product Code: FOX30 CT0119BLot Numbers: 1689092 & 1689093093ARTG Number: 208273