Oxylog Transport Ventilator 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Draeger Medical Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2015-RN-01256-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2015-12-23
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Some cases have been reported to draeger in which the "poti unplugged" error message was generated. in those cases an audible and visual alarm is generated, the breathing system releases pressure and the ventilation function stops operating. the error message is caused by increased electrical contact resistances of the controllers (setting potentiometers). the increased resistance is caused by an oxide layer that has formed on the controller over a lengthy period of time. no patient injury has been reported due to this issue.
  • 조치
    The oxide layer can only form if the controllers are rarely or never moved and according to product monitoring, some users never, or rarely, use the FiO2 controller. As a preventative measure to ensure the resistance taper of the controllers are sufficiently cleaned, Draeger is recommending the following addition to the daily device check: - Prior to turning on the Oxylog in readiness to conduct the daily device check, the staff member turns each of the controllers (Vt, RR, Pmax and FiO2) to the right and left stop positions (minimum and maximum values) at least ten times each. - The Oxylog can then be turned on and the onboard device check completed.