Patient Data Module (PDM) used with B850, B650, B450, Solar 8000i, MacLab / ComboLab (Multifunction patient monitoring)All PDMs produced from April 2007 through June 2010 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 GE Healthcare Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2013-RN-01255-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2013-11-27
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Fluid can pass through the seals in the top cover and come into contact with energised contacts in the circuit board. this can result in heat buildup, smoke, and discoloration. this heat buildup could potentially result in thermal injury to a patient or a caregiver.
  • 조치
    GE Healthcare is advising users to inspect the top cover for signs of heat buildup or discolouration. Remove from use if any evidence of damage is observed. GE is advising users to refrain from spraying fluid onto device. A new cover will be installed on all affected devices as a permanent fix when available. This action has been closed-out on 11/02/2016.