PenAdapt(an aspiration sheath that fits over an electrosurgical pencil body and leaves the tip or blade exposed) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Stryker Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2015-RN-00422-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2015-05-18
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    During packaging verification testing, a breach in the sterile barrier was observed in some of the units tested. additional testing identified microscopic aluminium metal shavings and foreign particles estimated in size to be between 0.01inches x 0.015 inches to 0.03 inches x 0.04 inches in some of the units tested. it was determined that these particles were introduced during the manufacturing process and were sterilised within the tyvek pouch with the product. the potential hazards of this product issue include patient exposure to infectious material, potentially leading to patient infection, and patient exposure to sterile foreign material potentially leading to foreign body inflammation.
  • 조치
    Customers are asked to check any PenAdapt items in their possession and quarantine any affected units immediately. A Stryker Representative will contact customers to coordinate an inspection of the affected items in quarantine. Any affected items will be removed by the Stryker Representative and a credit will be issued to the customer account. Stryker is also requesting the customers to review the patient records for issues that may have been associated with the use of the affected device, such as infection and/or inflammation.

Device

  • 모델명 / 제조번호(시리얼번호)
    PenAdapt(an aspiration sheath that fits over an electrosurgical pencil body and leaves the tip or blade exposed)Item number: 0702045027Multiple lot numbersARTG Number: 171568
  • Manufacturer

Manufacturer

  • 제조사 모회사 (2017)
  • Source
    DHTGA