PentaRay Nav Catheter (used for multiple electrode electrophysiological mapping of cardiac structures in the heart) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Johnson & Johnson Medical Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2014-RN-00519-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2014-05-09
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Biosense webster, a division of johnson and johnson medical, has identified an issue in the production process of the pentaray nav catheter that can lead to an insufficient bond of the distal tip to the catheter shaft. to date, with regard to the affected products, biosense webster has not received any reports of: (i) a complete tip separation from the shaft; or (ii) patient injuries or adverse events as a result of a partial tip separation from the shaft. there is no concern for patients who have already been successfully mapped with the affected products. the incidence rate of the tip to shaft separation is 0.11 % . partial separation exposes the internal components of the catheter to intracardiac blood and could lead to thrombus formation and could become a potential embolic source.
  • 조치
    Customers are advised to immediately check their inventory to quarantine if there are affected products.

Device

  • 모델명 / 제조번호(시리얼번호)
    PentaRay Nav Catheter (used for multiple electrode electrophysiological mapping of cardiac structures in the heart)Product codes: D128201, D128202, D128204, & D128205All lot numbersARTG number: 203362
  • 의료기기 분류등급
  • Manufacturer

Manufacturer