Phadia 1000 Instrument. An in vitro diagnostic medical device (IVD) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Abacus ALS Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2017-RN-01072-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2017-08-30
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    An error handling issue for code 7-101, liquid sensor error, has been reported on the phadia 1000 instrument. when error 7-101 occurs, immunocap dispensing will stop, whereas already dispensed samples will continue to be processed. due to an isw issue, the “retry” command, used to clear this error, does not function properly and there will be no further pumping of wash and rinse solution during the processing of the dispensed samples.Thus, if the error code 7-101occurs during the assay run, there may be a shortage of wash and rinse solution that will affect assay performance and test results if the operator chooses to respond to the error with the “retry” command.
  • 조치
    A mandatory ISW update will be provided to correct the problem. In the meantime, Abacus ALS advise you to take the following actions: - All instrument operators are to be instructed that if the 7-101 error occurs during an assay run, the “Stop” command must be selected. This will abort the run and results flagged as erroneous. Results from the assay run should not be approved. - Review records to determine if error 7-101 has occurred. If the instrument error message “7-101 LIQIUD SENSOR ERROR (Wash upper ON lower OFF)” has been reported. Abacus ALS will be contacting you to assist with this. - Fill in and return the supplied acknowledgement form - Phadia AB will have a revision of the current ISW version available soon. Technical Support staff will contact users in regards to the scheduling of this mandatory software installation.

Device

  • 모델명 / 제조번호(시리얼번호)
    Phadia 1000 Instrument. An in vitro diagnostic medical device (IVD)Article Number: 12-3800-01All instrument software (ISW) versionsAll Lot/Serial NumbersARTG Number: 192561
  • Manufacturer

Manufacturer