Philips Anaesthesia Machines IntelliSave AX700, Dameca MRI 508, Siesta i Whispa, Siesta iTS & Siesta Breasy 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Proact Medical Systems (Australia) Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2015-RN-00127-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2015-02-23
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    If the patient is connected to the anaesthesia machine in a volume controlled ventilation mode, and attempts a large spontaneous breath, it is possible for the patient to create a high negative airway pressure, around the same time as a volume controlled breath (mandatory breath).This pressure will be limited by the mechanical negative pressure limiter (npl) valve. the npl valve opens if the patient generates a negative pressure of -5 and -7.5 cmh20 (hpa). this opening pressure of the npl valve is defined at an inspiratory flow rate of 3.5-4.5 l/min. if the patient creates a higher flow rate than 4.5 l/min, the npl valve will continue to function but due to the resistance of the flow and time taken to reach the patient, the airway pressure may decrease further.
  • 조치
    Device Technologies is providing an updated addendum (Part Number 453564509701 56 052014) to the IFU and recommendation that "If the patient is breathing spontaneously, support ventilation modes (VSV, PSV) could be considered". This action has been closed-out on 08/08/2016.

Device

  • 모델명 / 제조번호(시리얼번호)
    Philips Anaesthesia Machines IntelliSave AX700, Dameca MRI 508, Siesta i Whispa, Siesta iTS & Siesta BreasyIntelliSave AX700Product Numbers: 866205/10623-00Dameca MRI 508Product Numbers: 866203/10651MRI-00Siesta i WhispaProduct Numbers: 866202/10651-00Siesta iTS Product Numbers: 866163/10653-00Siesta BreasyProduct Numbers: 866204/10652-00All serial numbers affectedARTG Number: 176848