Philips Anaesthesia Machines Siesta i Breasy, Siesta i Whispa, Dameca MRI 508, Siesta i TS & IntelliSave AX700 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Proact Medical Systems (Australia) Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2015-RN-00126-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2015-02-24
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    The spring-function of the control dial may not work, which may cause the control dial to get stuck when pushed, thus preventing its normal and expected function. it can be brought up to working condition again by simply pulling it out. the control dial is part of the user interface and serves an important role in setting and controlling the anaesthesia machine functions as it is used to adjust and confirm settings made by the user. for siesta i breasy, siesta i whispa and mri 508 freshgas regulation and ventilator mode change (manual or ventilator) can still be done as this is mechanical and/or soft key controlled and doesn’t need confirmation by the control dial.For siesta i ts and intellisave ax700 it is not possible to regulate freshgas delivery or change ventilator mode as this needs confirmation by the control dial.For all machine types ventilation will continue at the previously set parameters.
  • 조치
    Device Technologies is requesting all its customers to check their machines for the faulty control dial and is replacing control dials on devices exhibiting the issue. End users are provided work around instructions to assist in the continued use of affected units. The anesthesia machine can continue to be used as the control dial can be pulled out, making the control dial work. This action has been closed-out on 08/08/2016.

Device

  • 모델명 / 제조번호(시리얼번호)
    Philips Anaesthesia Machines Siesta i Breasy, Siesta i Whispa, Dameca MRI 508, Siesta i TS & IntelliSave AX700Siesta i BreasyProduct Numbers: 866204/10652-00Siesta i WhispaProduct Numbers: 866202/10651-00Dameca MRI 508 Product Numbers: 866203/10651MRI-00Siesta i TS Product Numbers: 866163/10653-00IntelliSave AX700 Product Numbers: 866205/10623-00Multiple Serial Numbers affectedARTG Number: 176848
  • Manufacturer