Philips HeartStart MRx Monitor/Defibrillator 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Philips Electronics Australia Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2014-RN-01275-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2014-12-05
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Issue 1: the mrx can be susceptible to interference from electrical fast transients (efts) when connected to ac or dc power, operating with a lan cable, or operating near a source of eft interference, which could cause therapy to be delayed or delivered inadvertently.Issue 2: if a user undertakes certain atypical clinical workflows (ie, when using external paddles for defibrillation or when using the periodic clinical data transmission (pcdt) option on the mrx), the mrx can exhibit unexpected behaviour. these workflows do not correspond to instructions in the mrx instructions for use (ifu) and are not expected to be performed by trained clinicians. in addition, these device behaviours have only been observed during internal testing, and have not been reported during clinical use. issue 3: the mrx could stop demand mode pacing due to an ecg leads-off condition when electrode-to-skin contact impedance values are outside design ranges for detection.
  • 조치
    A software upgrade will be implemented for all units affected by one or more of the above issues. Users are advised that they can continue to use their MRx prior to receiving the software upgrade, provided they follow the temporary workaround instructions provided in the customer letter.