Philips M1783A and M5526A 12 Pin Sync Cables 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Philips Electronics Australia Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2014-RN-01291-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2014-12-10
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    The following problems were found during internal testing:if electrical fast transits (eft) noise is injected into the monitor/defibrillator via ac mains power, and the monitor/defibrillator is connected to and receiving ecg signals from an auxiliary bedside monitor with an m1783a/m5526a 12 pin sync cable (8/24 feet) via the monitor’s ecg out port, then:1. when using a sync cable with the heartstart mrx or heartstart xl, eft noise can be mistaken as an r-wave. if this occurs when performing synchronized cardioversion, there is a potential for inducing ventricular fibrillation if shocks are synchronized to eft noise instead of the patient’s actual r-wave. 2. when using a sync cable with the heartstart xl+, eft noise can disable ecg monitoring, and potentially interrupt demand mode pacing, leading to a possible delay in therapy. a power cycle is required to resume ecg monitoring, however, fixed mode pacing is not impacted by this issue.
  • 조치
    Philips will replace all affected sync cables with a new sync cable free of charge. A Philips Healthcare representative will contact customers with affected devices to arrange for a replacement. This action has been closed-out on 05/12/2016.