PHILIPS MX 16-slice CT System, software: v1.1.3.1167; v1.1.4.21425; v1.1.4.21424; v1.1.4.21422 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Philips Electronics Australia Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2014-RN-01105-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2014-10-10
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    The orientation of images is displayed incorrectly. when operator chooses “gantry on the right side” from operation console, and the patient position is hfdl, hfdr, ffdl or ffdr, the patient position picture and text is not consistent. "l" and "r" of scan images are reversed and this is not consistent with the patient actual position. this issue may result in potential misdiagnosis or incorrect anatomy classification leading to unnecessary treatment/procedure.
  • 조치
    Philips is providing temporary work around instructions in the customer letter until updated software in a Service Pack is installed in the affected systems. The Service Pack will also correct the following other issues: - When a surview which is longer than 557mm is stopped by using “Abort” function during the scan process, the position of scan box in surview image is not correct. - For dental applications, when a patient scan is taken with Gantry tilted, and the tilt angle falls in the range from 1 to -1 degree (not including 0 degree position), the data post-processing may not operate normally and the images may not be displayed. - If the user stops a surview using the “Abort” function during the second part of the scan, the positions of the scan box are not consistent. This action has been closed-out on 09/08/2016.

Device

  • 모델명 / 제조번호(시리얼번호)
    PHILIPS MX 16-slice CT System, software: v1.1.3.1167; v1.1.4.21425; v1.1.4.21424; v1.1.4.21422ARTG number: 98868