Philips Network Firewall 867098 (Cisco ASA 5506 Firewall) used with the Philips IntelliVue Information Center iX or Classic Information Center 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Philips Electronics Australia Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2017-RN-01332-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2017-10-18
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Philips has identified a defect involving the clock signal component within the philips network firewall 867098 (cisco asa 5506 firewall). this component has a high probability of failing in appliances that have been running for greater than 18 months. where the clock signal component fails, the firewall will stop functioning and it will not boot and it will be unable to recover from the component failure. this failure will result in a loss of communication between devices that are separated by the firewall, which may cause the information center to reboot.
  • 조치
    Philips representatives will contact affected customers to arrange for replacement of all affected units supplied by Philips.