Philips Xper Flex Cardio Patient Monitoring Systems Revisions A and B 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Philips Electronics Australia Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2018-RN-00643-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2018-06-05
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Philips have identified an intermittent communication issue between the host system and the flex cardio (fc). it is possible for any of the following conditions to occur, which could result in delays in diagnosis or treatment:- boom monitor may not display all active waveform and/or vital sign data- delayed audible and visual alarms at the fc and boom monitor respectively, due to a mismatch between the alarm limits on the host system and the fc/boom monitor- after patient admission and case initiation at the host system, the case may not be fully transmitted to the fc. in this case, the boom monitor in the exam/procedure room will display vital sign data, but the fc will not provide audible alarms- after patient discharge and case termination at the host system, case termination may not be transmitted to the fc. in this case, the boom monitor in the exam/procedure room will continue to display vital sign data, and the fc will continue to provide audible alarms.
  • 조치
    Philips are advising the following to reduce the probability of intermittent communication issues: - The Flex Cardio must communicate with its host system via an isolated, dedicated Ethernet port - Do not turn the Flex Cardio device on and/or off while a patient case is open - Verify that the patient name is visible on the Boom Monitor, which is an indication that the patient case has fully transferred and audible alarms are active. Also, if all expected vital signs are not displayed on the Boom Monitor, close and then re-open the case to resynchronize the Flex Cardio and the host system. An addendum to the device’s Instructions for Use will be provided to all affected customers by a Philips representative. Philips will also review the contents of the Instructions for Use addendum with each customer and confirm understanding of the addendum’s contents. Affected customers will also be provided with an updated Software and Documentation Media Kit, which contains the IFU addendum.

Device

  • 모델명 / 제조번호(시리얼번호)
    Philips Xper Flex Cardio Patient Monitoring SystemsRevisions A and BAffected devices have at least one of the following numbers on its back: Service # 453564241901 and 453564483321ARTG Number: 229631(Philips Electronics Australia - Patient monitor system)