PLEGIOX Cardioplegia Heat ExchangerAll Standalone PLEGIOX Cardioplegia Heat Exchangers and those included in tubing sets. 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Getinge Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2017-RN-00978-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2017-08-08
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    The plegiox heat exchanger is used to set and maintain the temperature (for given flow rates and within the given temperature range) of blood cardioplegia and / or crystalloid cardioplegia solutions during extracorporeal circulation. the plegiox heat exchanger consists of two liquid circulations strictly separated from each other by a separating barrier to separate both liquids from each other and to effect thermal transfer at the same time; the design of the plegiox uses hollow fibres made of polyurethane. hollow fibres serve as barrier separating the circulating liquids while effecting thermal transfer. maquet cardiopulmonary gmbh (mcp) internal investigations have revealed that the specification of the performance factor (pf) at flow rates of 1 litre per minute is not always maintained as defined in instruction for use (ifu). deviations up to minus 20% were identified during in-house testing. zero adverse events have been reported in relation to this insufficiency.
  • 조치
    Getinge seeks to notify users that: When using the PLEGIOX, the flow rate should be no higher than 500 ml/min to ensure reasonable performance.

Device

  • 모델명 / 제조번호(시리얼번호)
    PLEGIOX Cardioplegia Heat ExchangerAll Standalone PLEGIOX Cardioplegia Heat Exchangers and those included in tubing sets.Reference Numbers: 703006246, 703006285, 703006317ARTG Number: 158781
  • Manufacturer

Manufacturer