Prodisc-C Vivo Cervical Disc Prosthesis (part of the Prodisc-C Vivo Cervical Disc Prosthesis System) Prodisc-C Vivo, uncemented convex, LD, 17x16mm, H5mm 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Johnson & Johnson Medical Pty Ltd t/a DePuy Synthes 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-00849-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2016-06-24
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Prodisc-c vivo implants are packaged sterile in a double barrier setup, with an outer foil package and an inner blister package. for the impacted lot, the tyvek lid may be missing from the inner blister package. manufacturer’s assessment has identified that due to the double barrier sterile packaging, the sterility of the implants will continue to be maintained when the tyvek lid is missing. the packaging non-conformance has the potential to result in intraoperative surgical delay. when the outer foil pouch package is opened, the implant may dislodge from the uncovered blister package and has the potential to fall outside of the sterile field.
  • 조치
    Users are advised to inspect stock and quarantine any remaining product for return to Johnson and Johnson. Patients who are implanted with the affected stock should be monitored in accordance with standard practice for the Prodisc-C Vivo treatment process.

Device

  • 모델명 / 제조번호(시리얼번호)
    Prodisc-C Vivo Cervical Disc Prosthesis (part of the Prodisc-C Vivo Cervical Disc Prosthesis System)Prodisc-C Vivo, uncemented convex, LD, 17x16mm, H5mmPart Number: 04.670.955SLot Number: 9736836ARTG Number: 133399
  • Manufacturer

Manufacturer