PROFEMUR NECK VAR/VAL 8DG, Long Cobalt Chrome, Part number PHAC1254(Modular neck component of the femoral head/stem prosthesis) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Wright Medical Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2015-RN-00737-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2015-08-19
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Microport orthopaedics inc. has identified unanticipated postoperative implant fractures related to this specific part number. these fractures are extremely rare, yet there have been 27 reports (failure rate of 0.275%) of fracture of the phac1254 component. all fractures involve part number phac1254 (long 8-degree varus) which sees the highest bending force due to its length and therefore represents the implant in this product family with the lowest mechanical strength under clinical conditions. no implant fractures have been reported in other pro femur cocr modular neck models. in the event the modular neck fractures, the patient may experience sudden pain, instability and difficulty walking/performing common tasks.
  • 조치
    Surgeons are advised that there is no evidence fractures can be anticipated by visual inspection or during surgery. When the modular neck fractures, the patient may experience sudden pain, instability and difficulty walking/performing common tasks. A break will require revision surgery to correct. Should a patient present with sudden pain, instability and difficulty walking/performing common tasks a postoperative fracture can be confirmed via x-ray. No additional patient follow-up is required however surgeons are advised to discuss the issues with their patients . This action has been closed-out on 04/05/2017.

Device

  • 모델명 / 제조번호(시리얼번호)
    PROFEMUR NECK VAR/VAL 8DG, Long Cobalt Chrome, Part number PHAC1254(Modular neck component of the femoral head/stem prosthesis) All lot numbersARTG number: 234443
  • Manufacturer

Manufacturer