Progressa Bed System with European (OIML) Version ScaleAll Progressa Beds manufactured prior to 16 May 2016 containing European (OIML) Version Scale 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Hill-Rom Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-00941-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2016-07-22
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Hill-rom has become aware of a potential software issue with some progressa beds equipped with the european (oiml) version scale. the issue affects how the air system utilises data from the scale to set bladder pressure points. if the software does not convert the patient weight to the air system correctly, then in some cases the pressure setting of the bladders may be higher or lower than the optimal pressure range. in some cases, patients that are already high risk may become more susceptible to pressure ulcers. this issue does not affect the accuracy or calibration of the progressa scale weighing function. the likelihood of occurrence is less than 57/100,000 units.
  • 조치
    Hill-Rom will be providing an upgrade to the software of the affected units. The upgrade allows for improved scale communications with the surface and system software, and will not affect how the scale and bed are operated. In the interim, Progressa beds equipped with the European (OIML) Version Scale may continue to be used while maintaining proper patient positioning practices.

Device

  • 모델명 / 제조번호(시리얼번호)
    Progressa Bed System with European (OIML) Version ScaleAll Progressa Beds manufactured prior to 16 May 2016 containing European (OIML) Version ScaleModel Number: P7500Serial Numbers: S1 12234946, S1 11791226 and O277AW0635ARTG Number: 221346
  • Manufacturer

Manufacturer