QMS Everolimus Immunoassay (used for the therapeutic monitoring of immunosuppressant drugs). An in vitro diagnostic medical device (IVD) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Thermo Fisher Scientific 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-01084-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2016-08-19
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    The manufacturer has determined that the affected lot may yield over-quantitation in patient samples. the manufacturer has conducted an internal assessment of product performance, and determined that over-recovery of whole blood assay results in the sub-therapeutic range by as much as 50% (for example, a reported assay result of 3.75 ng/ml instead of the actual whole blood concentration of 2.5 ng/ml). given standardised medication toxicity and close monitoring of patients for markers of efficacy, the risk of developing serious or long-range adverse health consequences due to a single reported erroneous everolimus concentration result is deemed to be negligible.
  • 조치
    Thermofisher is advising users to discontinue use and destroy any remaining stocks of the affected batch. A credit note will be issued for the affected product. A review of previously generated results should be considered by the laboratory director and treating physician.

Device

  • 모델명 / 제조번호(시리얼번호)
    QMS Everolimus Immunoassay (used for the therapeutic monitoring of immunosuppressant drugs). An in vitro diagnostic medical device (IVD)Catalogue Number: 0373852Lot Number: 72258007Expiry Date: 31 July 2017ARTG Number 235822
  • Manufacturer

Manufacturer