Radiometer 905-873 Fixation Kit N20 (used with Radiometer TCM CombiM and Philips TC Gas10 for continuous transcutaneous monitoring of carbon dioxide (tcpCO2) and oxygen (tcpO2) partial pressures) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Radiometer Pacific Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2015-RN-00636-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2015-07-15
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    The fixation kit consists of an adhesive which is the interface to the patient’s skin and on the other side, a plastic ring with a sensor retainer mechanism, which allows fastening a tc sensor 84 and tc sensor 54. in some cases, when the operator detaches the fixation ring from the paper liner, fractions of the paper liner’s surface coating sticks to the adhesive surface of the ring. the result is a varying degree of adhesiveness of the ring to the skin. if the ring and sensor fall off, the displaying data on the monitor will not be in range and thus identified by the medical professional. if the sensor is only in partial contact with the patient’s skin it may result in an incorrect measurement.
  • 조치
    Customers are advised to check inventory and stock in use for stock of the affected stock and discard them. Radiometer will provide free of charge replacement stock for stock that has been discarded. Users are advised to review customer results where affected stock was used and results were abnormal. This action has been closed-out on 15/08/2016.

Device

  • 모델명 / 제조번호(시리얼번호)
    Radiometer 905-873 Fixation Kit N20 (used with Radiometer TCM CombiM and Philips TC Gas10 for continuous transcutaneous monitoring of carbon dioxide (tcpCO2) and oxygen (tcpO2) partial pressures)Lot Numbers: 407221, 407241, 407251, 408201, 409231, 410021, 410141, 410281, 411171, 411201, 411261, 412111, 412201, 501081, 501131, 501191, 501311, 502031, 502131, 502171, 502231, 502261, 503091, 503231, 503311, 504081, 504111, 505041, 505141.ARTG Number: 165940