Radiometer ABL700 and ABL800 series blood gas analysers. An in vitro diagnostic medical device (IVD). Radiometer pH electrode, 945-614 with pCO2 membrane, 942-063 (D788) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Radiometer Pacific Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-01538-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2016-12-06
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    In july 2013 customers were informed about the possible risk of negative bias on some pco2 results on both blood and qc measurements. as a temporary countermeasure, users were required to manually narrow the qc ranges for pco2 and also perform extra qc measurements when replacing the membrane. the root cause investigation for the pco2 issue revealed a problem with gas leak at the ph electrode. in nov. 2014 a new improved blue sealing ring for the ph electrode was released. radiometer advised that the temporary countermeasure with the qc could be revoked upon installation of the improved sealing ring. further investigation has identified that over time, the dimensions of some of these new blue sealing rings changed. the blue sealing ring was improved and the updated version released in dec. 2015. tests were performed to monitor their performance over time and have shown that the dimensions of the updated sealing rings do not alter. these blue sealing rings are to be installed.
  • 조치
    Radiometer will replace the current sealing ring at the Planned Maintenance for the analyser(s) within the next 12 months. If users are experiencing any problems with pCO2 results, they are to contact their Radiometer representative to arrange an earlier time for the replacement of the sealing ring for the pH electrode. When the sealing ring is updated Radiometer will also replace any stock currently held of the 834-682 sealing ring. A Service report will be provided when the sealing ring is replaced.

Device

  • 모델명 / 제조번호(시리얼번호)
    Radiometer ABL700 and ABL800 series blood gas analysers. An in vitro diagnostic medical device (IVD). Radiometer pH electrode, 945-614 with pCO2 membrane, 942-063 (D788)ARTG Number: 229300