Radiometer ABL800 blood gas analysers with software versions 6.14 through to Version 6.18. An in vitro diagnostic medical device (IVD). 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Radiometer Pacific Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2018-RN-00926-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2018-07-09
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Radiometer has recently become aware that the abl800 series analysers could potentially induce a patient mix-up when “confirmation of demographics” is enabled as part of the query to aqure, radiance, or a his lis system. please note that the default setting for this function is ‘disabled’.The described error may potentially impact upon patient management as the physician may subject the patient to unnecessary treatment or erroneously withhold necessary treatment from the patient based upon the mixed-up of patient identification. in a worst-case scenario, where a critical parameter, such as k+ or po2, is interrogated, based upon a false clinical diagnosis inferred from the erroneous measurement, a critically ill patient may be subjected to lack of critical treatment.
  • 조치
    Radiometer is advising that they are currently working on revising the software to correct this issue. Once release, a Radiometer representative will contact users to schedule a visit for the software upgrade. In the interim, users are requested to confirm if "Confirmation of Demographics" is disabled on their analysers. If disabled, the issue will not be experienced. If enabled, one of the two following counter-measures (CM) must be implemented: CM #1: Disable “Confirmation of Demographics” as follows: a. Logon to the analyzer. b. Select “Menu” > “Utilities” > “Setup” > “General Setup” > “Communications” > “Automatic Data Request” c. Remove the checkmark for “Confirm requested data”, if present. d. Select “Close”. Or CM #2: If disabling the “Confirmation of Demographics” (CM #1) is not possible due to e.g. internal procedures, instruct all operators of the analyser(s) to not introduce a new sample when the menu for “Confirmation of Demographics” is being displayed.

Device

  • 모델명 / 제조번호(시리얼번호)
    Radiometer ABL800 blood gas analysers with software versions 6.14 through to Version 6.18. An in vitro diagnostic medical device (IVD). ARTG Number: 228382(Radiometer Pacific Pty Ltd - Instrument/analyser IVDs)