Radiometer ABL800 series with FLEXQ module. An in vitro diagnostic medical device (IVD) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Radiometer Pacific Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2017-RN-01316-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2017-10-16
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Radiometer has identified that the registration receipt received after placing a sampler in the flexq can be associated with a different sampler if the barcode quality on the receipt has been affected. this issue may occur in the following situation:1. the analyser is setup to use registration receipts providing a barcode to identify the sampler and thereby the patient result;2. the sampler is placed in the flexq and a registration receipt is printed from the analyser;3. the quality of the barcode on the receipt is poor due to improper handling/damage or poor print quality;4.The registration receipt is later scanned on the analyser to print the result and the analyser misinterprets the barcode; and/or5. in the event that the misinterpreted barcode matches that of a previous registration receipt printed on that analyser, the damaged registration receipt may be associated with the previous sampler.
  • 조치
    Radiometer is advising users who are testing a sampler in the FLEXQ and printing a registration receipt to verify that the patient results and demographic information match. Radiometer is developing a software upgrade to change the registration receipt barcode to a more robust version to remove the possibility of the barcode being misinterpreted. Radiometer will notify users when the upgrade is available.