Radiometer TCM4 series analyser base units (Transcutaneous and non invasive blood gas analysers) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Radiometer Pacific Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2012-RN-01207-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2012-12-07
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Radiometer has become aware that specified tcm4 series base units do not comply with the medical safety standard iec 60601-1, 2nd edition, subclause 15(b).The implication is that there is a potential risk of receiving an electric shock under certain conditions as follows:the monitor is connected to the mains outlet and the mains switch is on; the battery is fully charged (or no battery is present); the operator disconnects the line cord (at either end) without first switching the mains switch on the back of the monitor to off; the operator touches the two metal pins on the mains plug or on the monitor's line cord receptacle after disconnecting the line cord.
  • 조치
    Radiometer is informing users of the issue and is providing fluorescent warning stickers to be placed on the device and work around instructions to mitigate the risk if the device must be used. A hardware correction is forthcoming to permanently correct the problem.