Radiometer TCM5 FLEX / BASIC Transcutaneous Monitor(when used with ECG or EEG monitoring) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Radiometer Pacific Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2017-RN-01327-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2017-10-18
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Radiometer has recently received reports of occurrences of the tcm5 causing interferences on ecg and eeg monitors. there is a possibility that tcm5 flex / basic can cause interferences to ecg or eeg measurements, however, only when applied on the same patient.For this reason the instructions for use include the following warning:warning – risk of interferenceother instruments sensitive to electromagnetic interference may be affected by the tc measurement. if this is the case, contact authorised service personnel.The described effect may potentially lead to incorrect measurement results by 3rd party devices, such as an ecg monitor due to interference by the tcm5 flex / basic monitor. these incorrect measurement results may potentially lead to incorrect treatment or management of the patient.
  • 조치
    Radiometer is advising users who perform ECG measurements in combination with the TCM5 FLEX/ BASIC on a patient to implement one of the counter measures below to reduce/prevent the interference. 1. Disconnect the TCM5 FLEX/BASIC from the mains supply and run it on battery only. This will prevent the interference from occurring on an ECG or EEG monitor. 2. Do not use both the TCM5 FLEX/BASIC and an ECG or EEG monitor simultaneously on the same patient. Radiometer is revising the Instructions for Use and will provide these to customers when available. Radiometer will contact affected customers when a solution becomes available.

Device

  • 모델명 / 제조번호(시리얼번호)
    Radiometer TCM5 FLEX / BASIC Transcutaneous Monitor(when used with ECG or EEG monitoring)All Serial NumbersARTG Number: 140333