RaySearch RayStation 3.0 and 3.5 (software versions 3.0.0.251, 3.5.0.16, 3.5.1.6) (Radiation treatment planning and analysis software) Date Manufactured: 15/02/2013 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Emergo Asia Pacific Pty Ltd T/a Emergo Australia 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2013-RN-00734-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2013-07-19
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Under certain conditions resampling of regions of interest (roi) geometries to the dose grid representation is not correctly calculated, and an additional voxel layer is added incorrectly on the superior side of each roi. the problem is more pronounced for very small roi's, leading to over-estimation of dose maximum and under-estimation of the minimum dose.The above problem also affects dose calculated with material overrides, however this should cause no noticeable impact.
  • 조치
    Emergo are providing work around instructions for users to follow. The issue will be corrected in RayStation software version 4.0 which is due for release in July 2013.