RaySearch RayStation 5 (Radiation therapy treatment planning system) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Emergo Asia Pacific Pty Ltd T/a Emergo Australia 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-00717-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2016-06-08
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    This notice concerns an issue found with the treatment plan report in raystation 5. for a treatment plan with multiple beam sets, the table for region of interest (roi) plan dose statistics in the report may show the statistics for a beam set dose instead of the plan dose. in the case where a clinical decision is based on the erroneous roi plan dose statistics, a patient could potentially receive a higher than acceptable radiation dose. the error can only be triggered when using a report template where statistics for a beam set dose is included as the last dose statistics prior to the plan dose statistics. for such a report template, all reports for multiple beam set plans would always be affected.
  • 조치
    Users are advised to inspect all report templates that are intended to be used for plan with multiple beam sets, and to ensure that "POI Dose Statistics (Plan)" is included immediately before "ROI Dose statistics [Plan dose]" in the report content area. A software upgrade will be implemented as a permanent correction. This action has been closed-out on 16/02/2017.

Device

  • 모델명 / 제조번호(시리얼번호)
    RaySearch RayStation 5 (Radiation therapy treatment planning system) Affected software versions: 5.0.0.37, 5.0.1.11, 5.0.2.35 ARTG Number: 195288
  • Manufacturer

Manufacturer