RayStation 3.0, 3.5, 4.0, 4.5, 4.7, and 5(Radiation therapy treatment planning system) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Emergo Asia Pacific Pty Ltd T/a Emergo Australia 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-00904-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2016-07-12
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    There is an issue with the evaluation of biological clinical goals in raystation 3.0, 3.5, 4.0, 4.5, 4.7, and 5. this issue occurs when adaptive replanning (rayadaptive) is used in combination with biological models (raybiology) and can only occur for a user licensed for rayadaptive and raybiology concurrently. biological clinical goals for an adapted plan based on a new ct (not the planning ct used for the original plan) are evaluated using incorrect dose values and will therefore display incorrect values. the issue is present in all views displaying biological clinical goals, for adapted plans based on a new ct. the magnitude of the error is irregular and largely dependent on the differences between the original planning ct and the new ct. the computed tcp/ntcp values may be smaller or larger than the correct values. the issue has not caused any patient mistreatment or other incidents.
  • 조치
    The issue will be resolved in the next version of RayStation, scheduled for market release in December 2016. In the interim, users are advised: - Do not use the values displayed for biological clinical goals in clinical decision making when performing adaptive replanning. - The user is referred to using corresponding biological objective functions or biological functions defined in the Biological Evaluation module. These are evaluated correctly. - As stated in the RayStation Instructions for Use, always evaluate the physical dose.

Device