RayStation and RayPlan 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Emergo Asia Pacific Pty Ltd T/a Emergo Australia 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2017-RN-01436-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2017-11-23
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    It has been identified that an issue found with the electron dose calculation in raystation 3.5, raystation 4.0, raystation 4.5, raystation 4.7, raystation 5, raystation 6, rayplan 1 (raystation 4.9) and rayplan 2 (raystation 6). when calculating electron monte carlo dose with a very large number of histories, the dose calculation may be wrong.To the best of our knowledge, the issue has not caused any patient mistreatment or other incidents. however, the user must be aware of the following information to avoid incorrect dose calculations during beam modelling and treatment planning.Electron dose calculation in raystation uses a monte carlo dose engine, where a large number of particles are simulated. each simulated particle is called a history, adding to the simulated dose. the number of histories can be adjusted for each dose calculation.
  • 조치
    Operators should inspect their product and identify all installed units with the above software version number(s), then confirm you have read and understood this notice using the supplied Client Response Form. Operators should not use a number of histories above 6 million for electron dose calculation and should inspect all commissioned electron beam models for any applicator calculated with greater than 6 million histories. This issue will be resolved in the next version of RayStation, scheduled for market release in December 2017

Device

  • 모델명 / 제조번호(시리얼번호)
    RayStation and RayPlanModel or Catalogue: RayStation 3.5, RayStation 4.0, RayStation 4.5, RayStation 4.7, RayStation 5,RayStation 6, RayPlan 1 (RayStation 4.9) and RayPlan 2 (RayStation 6)Multiple Lot and Serial numbersARTG number:195288