RE51021 / MuSK-Ab ELISA 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Abacus ALS Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2017-RN-01435-1
  • 사례 위험등급
    Class III
  • 사례 시작날짜
    2017-11-23
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    This notice is to inform you about a recall action for ibl international product ibre51021 musk-ab elisa lot emu124 with expiry date 31-10-2018. this action has been undertaken by the australian distributor abacus als following consultation with the therapeutic goods administration (tga). no other lot numbers of this product are affected by this recall. description of the problem: the user of this assay may notice that the antiserum (#kemu521, lot: s882) and assay buffer (#kemu751, lot: s881) appears contaminated (cloudy and contain particles). this issue is caused by a contamination of lot # emu124.
  • 조치
    Users should inspect stocks and quarantine all units of IBRE51021 Lot EMU124, then complete and return the supplied Reply Form as directed for stock to be recovered and replaced.Complete the Reply Form even if you have no stock which is subject to recall, as Abacus ASL requires this information to reconcile this process. Retain the letter in a prominent position for one month in case stock is in transit. If any affected stock has been transferred from your Laboratory to another, immediately let that laboratory know of the recall and inform Abacus ASL.

Device

  • 모델명 / 제조번호(시리얼번호)
    RE51021 / MuSK-Ab ELISA Lot EMU124ARTG number: 210637