Ready-Prep %CDT by HPLC using the CDM software version MVC133091. An in vitro diagnostic medical device (IVD) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Bio-Rad Laboratories Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2017-RN-01532-1
  • 사례 위험등급
    Class III
  • 사례 시작날짜
    2017-12-20
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Bio-rad have observed that the lot numbers for mobile phase 2 [mp][2] and mobile phase 3 [mp][3] on the cd-rom cover print and within the method for the cdm software are not correct for the affetced cd-rom, due to a defect in the manufacturing process of the cd-rom.Thus, the printed numbers on the cd-rom do not correspond to the correct lot numbers of mobile phase 2 and mobile phase 3 which were delivered with the associated kits.The wrong lot numbers within the cdm method will affect only customers using the cdm software: the wrong lot numbers within the method of the cdm software will result in transfer of the wrong lot numbers into the users cdm software, and false lot numbers printed in the summary reports of the analytical runs. this may result in difficulties of the traceability of used mobile phases. due to the fact that the wrong lot numbers are not used for any other reagents or components of any %cdt by hplc test there is no risk of false traceability.
  • 조치
    Bio-Rad is advising users to immediately check the documentation obtained by using the CD-ROM of the affected lots. If users have used the CD-ROM to read in the lot numbers into the CDM software, correct the MP2 and 3 lot numbers in the CDM software manually (as described in the CDM manual Version 4.0, chapter 5.1.4). Alternatively, discontinue use of the CD-ROM and Bio-Rad will replace the CD-ROM when the quality issue is resolved.

Device

  • 모델명 / 제조번호(시리얼번호)
    Ready-Prep %CDT by HPLC using the CDM software version MVC133091. An in vitro diagnostic medical device (IVD)Catalogue Number: 1956670, 1956671Serial numbers: 64121977, 64139856, 64139684, 64142119, 64121979Expiry Date: 15-03-2018ARTG Number: 194835