Rechargable Lithium Ion Battery used in Beneview T1 Patient Monitors 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Mindray Medical Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2017-RN-01065-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2017-08-16
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Mindray has identified the potential for overheating of the rechargeable lithium ion battery used in beneview t1 patient monitors. the initial failure mode and clinical assessment indicated that the problem may result in the battery overheating, smoke, and the possible melting of the monitor shell under extreme circumstances. to date, there are no reports of patient injuries associated with this potential issue.
  • 조치
    1. Pass this Notice to all within your organisation or to any organisation where the potentially affected device(s) have been transferred. 2. Stop using the Rechargeable Lithium Ion battery of Beneview T1 Patient Monitor. Mindray service representatives will contact users to replace the affected batteries. 3. Store the affected batteries properly as follows; · Do not charge batteries. · Prevent short circuiting of the batteries. For example, covering the two ends of the connector with insulating tape. · Store the batteries at cool places(=21°C) away from combustibles. · Store the batteries separately, do not mix, overlay or pile up them with other batteries. 4. Mindray Service Representatives will visit to collect and dispose of the batteries.