Reflection Deadblow Mallet 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Smith & Nephew Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2017-RN-00301-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2017-03-01
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Smith & nephew has received a small number of complaints reporting cracks in the welds located on the head of the mallet, which is filled with lead beads. in the reported cases some of the lead beads and particles escaped from the mallet into the surgical wound. as these failures occurred intra-operatively and appropriate actions were taken to remove the lead beads from the wound during surgery, no further adverse effects were reported. smith & nephew is recalling the mallets as the risk of lead escaping from the mallet and the potential adverse effects associated with lead exposure were previously unidentified risks. based on the low occurrence (0.03%) of the worst case failure mode defined above and the high detectability should any lead beads escape into the surgical wound, no proactive follow-up is recommended.
  • 조치
    Smith & Nephew are advising customers to immediately locate and quarantine affected devices. Smith & Nephew will arrange for the return of the devices and the provision of unaffected replacement mallets.

Device

  • 모델명 / 제조번호(시리얼번호)
    Reflection Deadblow MalletModel Number: 71362106All Batch NumbersARTG Number: 104373
  • Manufacturer

Manufacturer