Remote controls for JUPITER, ARTIS and TruSystem 7500 operating tables 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Trumpf Med Aust Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2015-RN-00182-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2015-03-13
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Trumpf medical received reports that the operating tables sporadically moved independently and without the user deliberately pressing an operating element. no patients or users were injured. the investigations determined that the root cause was a button on a remote control which was mechanically damaged by external influences. as a result of the damaged button, specific vibrations or movements of the remote control can emit a valid movement command at any time, either continuously or sporadically.This action was undertaken prior to the notification to the tga.
  • 조치
    Trumpf Medical is providing instructions as a supplement to the user manual and requesting the users to check the remote controls for visual damages.Trumpf Medical is requesting the customers to avoid using the remotes with visual damages to the key pads.Trumpf Medical is arranging the upgrade of the existing remote controls with a redesigned membrane keypad to eliminate the known mode of failure. This action has been closed-out on 19/02/2016.

Device

  • 모델명 / 제조번호(시리얼번호)
    Remote controls for JUPITER, ARTIS and TruSystem 7500 operating tables
  • Manufacturer

Manufacturer