Renasys Transparent Film and Renasys Gauze Dressing Kits (used to dress, protect and seal wounds undergoing Negative Pressure Wound Therapy (NPWT)) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Smith & Nephew Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-01065-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2016-08-10
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Smith & nephew have identified seal integrity failures on the folded sterile renasys transparent film pouches during transit testing. all the seal integrity failures identified during the testing occurred at the point where the sterile renasys transparent film pouches were folded during the packaging process.The loss of sterility has the potential to contribute to a local infection. to date, smith & nephew is not aware of any complaints reported in relation to this issue that has led to a local infection.
  • 조치
    Smith & Nephew are advising users to inspect stock and quarantine all affected units. In the event that a Transparent Film is required for NPWT therapy, the responsible Health Care Professional should be consulted to determine an alternative Transparent Film or alternative treatment. Affected stock can be returned and replaced with unaffected product.

Device

  • 모델명 / 제조번호(시리얼번호)
    Renasys Transparent Film and Renasys Gauze Dressing Kits (used to dress, protect and seal wounds undergoing Negative Pressure Wound Therapy (NPWT))Part Numbers: 66800394, 66800932, 66800933, 66800934, 66800935, 66800936, 66801255, 66801256, 66801257, 66801258Multiple Products and Lot Numbers affectedARTG Number: 149671
  • Manufacturer

Manufacturer