Sciatic Nerve Retractors 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Johnson & Johnson Medical Pty Ltd t/a DePuy Synthes 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2018-RN-00334-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2018-04-18
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Johnson and johnson has identified the potential for micropores to form on the hollow handle of the sciatic nerve retractor. the pores may increase in size, allowing fluid to enter the hollow handle. discoloration and moisture were identified within the packaging of the affected lots of sciatic nerve retractors indicating that these pores may be present in the affected devices.Even with diligent reprocessing/sterilisation of the retractor, subsequent patients may be at risk for infection and adverse tissue reaction due to retained fluids leaking out during use.
  • 조치
    Johnson and Johnson are advising users to inspect stock and return any affected product as soon as possible.

Device

  • 모델명 / 제조번호(시리얼번호)
    Sciatic Nerve Retractors Part Numbers: 03.100.013, 03.100.014Lot Numbers: T140390, T140675, T140677 ,T145819, T140566, T140673, T151369ARTG Number: 296161Johnson & Johnson Medical t/a DePuy Synthes - Wound retractor, intermediate/deep