Scorpio Patella Clamp (used to hold the patella in place during the cementing process) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Stryker Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-01063-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2016-08-10
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Stryker orthopaedics has received four reports of disassociated components of the scorpio patella clamp. an investigation revealed that the press-fit specifications between the pins and either one or both clamping subcomponents were not met. no adverse patient consequences were reported. the instrument components, including pins and clamping subcomponents may potentially disassociate and fall into the wound intra-operatively, necessitating retrieval. as such, the potential harms may include: - complications associated with extended surgery time - inflammatory response - tissue damage - revision surgery to retrieve loose components.
  • 조치
    Stryker is advising users to inspect stock and quarantine the whole tray immediately if the affected product is found. A Stryker representative will coordinate a replacement clamp or the return of the whole tray that the affected item is in and provide access to unaffected trays as required. This action has been closed-out on 18/05/2017.

Device

  • 모델명 / 제조번호(시리얼번호)
    Scorpio Patella Clamp (used to hold the patella in place during the cementing process)Item Number: 3182-1000Tray Item Numbers: 31821000-T, 33023024-T, 80002017-T, 80106661-T, 8200TSX-TMultiple Lot Numbers affectedARTG Number: 140892
  • Manufacturer

Manufacturer

  • 제조사 모회사 (2017)
  • Source
    DHTGA