SENSE Body Coil 1.5T used with ACS NT, Intera 1.5T and Achieva 1.5T MR Systems Elevated temperature of the SENSE Body Coil 1.5T cables 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Philips Electronics Australia Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2015-RN-00584-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2015-06-30
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    The instructions for use state how (and how not) to position the sense body coil 1.5t and its cables in relation to the patient and mr bore in order to avoid a hazardous situation. when these instructions are not followed during an examination, the positioning of the coil cables can create unintended resonances via the patient or elevated coil cable temperature due to rf coupling to the qbc.
  • 조치
    Philips is reiterating the warnings in the IFU regarding positioning the cables and leads. A sleeve is being developed that will be applied to the cable on the SENSE Body Coil 1.5T and provide consistent spacing between the cable and the patient, thereby reducing the chance of RF interaction and heating.