SerenoCem Granules BCG050 (Orthopaedic bone cement) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Endotherapeutics Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-00439-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2016-07-12
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    The manufacturer, corinthian surgical ltd has been made aware of a cluster of cases in the united kingdom where it has been observed that there has been some bone re- absorption in the area adjacent to the granules in the middle ear. this can lead to exposure of the dura and facial nerve, potentially resulting in infection and/or brain herniation. as a precautionary measure corinthian surgical ltd is temporally suspending the supply of serenocem granules and there should be no further use of the granules at this time. the cause of the reported cases of bone resorption is currently under investigation.
  • 조치
    Endotherapeutics is recommending that if a patient presents for a routine checkup or at the 2 yearly follow up, to perform a CT scan. Outside of these circumstances, patients should undergo a CT scan at the clinical discretion of the treating surgeon. Any remaining unused stock should be quarantined and returned to Endotherapeutics. The investigation is currently ongoing, and further communications may be required following completion of the investigation and in consultation with the Therapeutic Goods Administration (TGA).

Device

  • 모델명 / 제조번호(시리얼번호)
    SerenoCem Granules BCG050 (Orthopaedic bone cement)All lots affectedARTG Number: 152441
  • Manufacturer

Manufacturer