SERVO-i/s ventilator systems with software version 8.00.00 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Maquet Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-00782-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2016-06-15
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Maquet have become aware of improper behaviour introduced in servo-i and servo-s ventilator system sw version 8.00.00. the improper behaviour in system sw version 8.00.00 appears as follows:· if ventilation is started with infant factory default settings, the following settings will apply: pressure control mode; and 10 cmh2o pc above peep.· the system will deliver a pressure of 20 cmh2o above peep as long as the factory default setting 10 cmh2o is used unchanged.For the infant patient category a delivered pressure of 20cmh2o above peep instead of preset 10cmh2o above peep may cause serious injury. the error was discovered at the production site and there are no reported injuries. the issue would be apparent in the pressure curve and peak pressure displayed on the screen.
  • 조치
    Maquet is advising users to change the default settings of pressure above PEEP in the infant patient category. The ventilator will only deliver a pressure of 20 cmH2O above PEEP if the default start-up configuration of the pressure has not been changed. A new system software version 8.00.01 is available that will eliminate this behaviour. Maquet has initiated an immediate update of all affected SERVO-i and SERVO-s ventilator systems. This action has been closed-out on 27/01/2017.

Device

  • 모델명 / 제조번호(시리얼번호)
    SERVO-i/s ventilator systems with software version 8.00.00Affected devices: SW Service Release v8.00.00 SERVO-I - 68 85 247 and SW Service Release v8.00.00 ERVO-s – 68 85 259ARTG Number: 162468
  • Manufacturer

Manufacturer