Siemens Somatom Emotion, Sensation, Perspective and Scope CT Systems with software versions VB42B, VC10A, VC20A, VC20B, VC28A, VC30A, VC30B and VC40A 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Siemens Healthcare Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2018-RN-00183-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2018-02-27
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Siemens healthcare became aware of possible incorrect tube current calculations by the care dose4d algorithm for scans that include the head, when based only on p.A. (posterior-anterior) or a.P. topograms. depending on the geometrical shape of the skull bone, it may happen in rare cases that the calculated dosedistribution is not appropriate and could lead to unnecessary radiation exposure. the described issue will not occur when using a lateral topogram instead of a p.A. or an a.P. topogram. there have been no injuries reported as a result of this issue.
  • 조치
    Siemens is strongly recommending using lateral topograms for all head scans. Considering that the p.a. or a.p. topogram is preferred in specific protocols such as in CarotidAngio scans and RT_ HeadNeckShoulder, it is strongly recommended to deactivate the CARE Dose4D function for this type of scan. For software versions VB42B, VC10A, VC20A, VC20B, VC28A, VC30A, VC30B, VC40A, to prevent any possible deterministic radiation effects on the patient's skin or eye lenses, Siemens Healthcare implemented a dose alert according to the technical standard IEC 60601-2-44. A warning will be shown and has to be confirmed by the user if the accumulated CTDIvol for the ongoing examination exceeds the alert threshold in any z-position. The default setting for the threshold is adjusted to 1000 mGy.