SIGMA Fixed Bearing Tibial Inserts - PLI, Cruciate Retaining, Posterior Stablised (used as part of SIGMA Knee System) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Johnson & Johnson Medical Pty Ltd T/A Depuy Australia 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2013-RN-00799-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2013-08-19
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    The compatibility chart on the product label and or reference chart literature/poster indicates that the size 6 posterior stabilized (ps) femoral component is compatible with the size 4 posterior stabilized (ps) tibial inserts and size 4 stabilized plus (sp) tibial inserts. the size 6 ps femoral component is not cleared for use with the size 4 posterior stabilized (ps) tibial inserts and size 4 stabilized plus (sp) tibial inserts, and should not be implanted with either of these size 4 insert.
  • 조치
    Existing OR Reference Charts on posters should be discarded and replaced with revised OR Reference Charts, which will be supplied to the customers by JJM sales representatives.

Device

  • 모델명 / 제조번호(시리얼번호)
    SIGMA Fixed Bearing Tibial Inserts - PLI, Cruciate Retaining, Posterior Stablised (used as part of SIGMA Knee System)Multiple model numbers affectedAll lots manufactured after July 2010ARTG Numbers: 138289, 98092
  • Manufacturer

Manufacturer