Single and dual LCD monitor suspensions used on Precision RXi fluoroscopic imaging systems 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 GE Healthcare Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2015-RN-00575-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2015-06-26
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Ge healthcare has become aware of a reported incident where a monitor fell from the suspension. a fall of a monitor could result in an injury to a patient or operator. there have been no injuries reported as a result of this issue.
  • 조치
    A GE Healthcare representative will contact affected customers to arrange for the correction. In the interim, users are advised to perform a visual check before and after each use for the presence of the four (4) mounting screws. Check the secureness of these screws by hand. If these screws are present and secure, users may continue use of the system. If any of these screws are missing or can easily be turned by hand, follow the precautions below before continuing use with the monitor(s). - Position the monitor suspension in the most frequently used position and limit further movement as much as possible. - Clearly inform (e.g. through signage and verbal instructions) the users and other auxiliary personnel who may come in contact with the system to not move the monitor suspension once in position. - Contact the GE Healthcare representative to inform them of any loose or missing screws. This action has been closed-out on 17/08/2016.