sleep.safe Set, sleep.safe Set Plus and sleep.safe Set Paed(Tubing sets for Peritoneal Dialysis) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Fresenius Medical Care Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2016-RN-01612-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2016-12-20
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    During the continuous monitoring of sleep safe tubing systems on the market, fresenius medical care has become aware of an increased number of leakages caused by overwelded coils on the patient and the drainage line. overwelded coils can lead to difficulties in unrolling the patient and drainage line and can result in rupturing the lines with pinholes, and thus leakages may occur. the possible damage might not be visually detected. the sleep safe cycler can also not detect such a leakage. therefore, there is a potential risk of infection for the patient, particularly the development of peritonitis. the problem of overwelded lines has been analysed and solved in subsequent production.
  • 조치
    Fresenius will be contacting patients (users) via hospitals and organising the exchange of potentially affected stock with replacements. In the interim, users are to follow the detailed handling instructions provided in the customer letter. Users are reminded to: - Unroll carefully the tube coils of the sleep safe sets. - If unrolling is not possible, discard the sets. - If unrolling is successful, check carefully the patient line for leakages during priming. For more information, please refer to https://www.tga.gov.au/alert/sleep-safe-tubing-systems-used-peritoneal-dialysis .

Device

  • 모델명 / 제조번호(시리얼번호)
    sleep.safe Set, sleep.safe Set Plus and sleep.safe Set Paed(Tubing sets for Peritoneal Dialysis)sleep.safe SetBatch Numbers: XAS131, XCS241sleep.safe Set PlusBatch Numbers: XBS011, XBS252, XBS253, XCS221, XES09100, XES10100, XES12100, XES17100, XES31100sleep.safe Set PaedBatch Numbers: XAS221, XBS043ARTG number: 123388