SoKINOX NO (Nitric Oxide) Delivery and Monitoring System 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Maquet Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2017-RN-00851-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2017-07-12
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Getinge has identified an issue where during no (nitric oxide) therapy, when a no cylinder runs nearly empty the system will indicate this by the "almost empty cylinder" icon. when the cylinder runs empty it will then trigger the alarm no cylinder 1 missing (or empty) or no cylinder 2 missing (or empty). this is indicated on the user interface by a crossed out cylinder icon. the system automatically shifts over to the other cylinder.Due to a pressure increase after the cylinder switch-over there are two possible issues identified:1. the cylinder status icon can change and erroneously indicate remaining gas in an empty no gas cylinder.2. the alarm “no cylinder 1 missing (or empty)” or “no cylinder 2 missing (or empty)” can be deactivatedif the empty no cylinder is not replaced, both no cylinders can run empty and cause a stop in no delivery. to date, there have been no reports of patient injury as a result of this issue.
  • 조치
    Getinge/Maquet will initiate an immediate update of all affected SoKINOX systems as soon as the system version 1.3 is released in Australia. Once version 1.3 available, customers will be contacted by a Getinge/Maquet sales or service representative to plan for the update of the device.

Device

  • 모델명 / 제조번호(시리얼번호)
    SoKINOX NO (Nitric Oxide) Delivery and Monitoring SystemDevice Number: 66 94 550ARTG Number: 273803
  • Manufacturer

Manufacturer