Solar 8000M and Solar 8000i Patient Monitor software version 5.4, 5.5 or 5.6 (multi-function patient monitor) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 GE Healthcare Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2014-RN-00114-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2014-02-05
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    The solar 8000m/i automatic view on alarm (avoa) feature will stop functioning if the patient monitor is discharged from the cic central station while a remote automatic bedside alarm view is currently being displayed on the solar. once this issue occurs, subsequent remote avoa alarms will not automatically appear on the left side of the solar display. in addition, the user interface menus on the discharged patient monitor will not function correctly. this may result in delayed treatment due to missed alarms that may be serious. this issue can be fixed by rebooting the solar 8000m/i.Parameter, waveforms and alarms associated with the patient directly connected to the solar 8000m/i patient monitor are not affected.
  • 조치
    GE Healthcare is advising users that if patients are discharged from the CIC central station to discontinue use of the AVOA feature and is providing instructions to disable the feature. If patients are not discharged from the CIC central station the AVOA feature can continue to be used. This action has been closed-out on 29/01/2016. GE Healthcare will be implementing a software upgrade for all Solar 8000i / M monitors.

Device

  • 모델명 / 제조번호(시리얼번호)
    Solar 8000M and Solar 8000i Patient Monitor software version 5.4, 5.5 or 5.6 (multi-function patient monitor)ARTG Number: 92726
  • Manufacturer

Manufacturer