SOMATOM CT System SOMATOM Definition Flash running software version syngo CT2012B SOMATOM Definition Edge running software version syngo CT2012B SOMATOM Definition AS running software version syngo CT2012B SOMATOM Definition running software version syngo CT2012B SOMATOM Definition AS running software version syngo CT2013 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Siemens Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2013-RN-01317-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2013-12-11
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Siemens has identified a software malfunction that may lead to a system display freeze during interventional procedures. if the function, "auto save last displayed images as key images" is activated during an interventional procedure, the software configuration allows conflicting access to the ct system image database. in such situations, a system display freeze might occur. normal interaction with the system is then no longer possible. no error message is issued and only a system restart resolves the problem. in case of a sudden system display freeze the interventionalist will potentially not immediately recognise the malfunction and continue the interventional procedure. there is a possibility of injury to a patient.
  • 조치
    Siemens advising users to deactivate the "Auto save last displayed images as key images" function to prevent the issue from occurring. Images can continue to be saved manually. Siemens is providing a software update to permanently correct the issue in February 2014. This action has been closed-out on 29/01/2016.

Device

  • 모델명 / 제조번호(시리얼번호)
    SOMATOM CT System SOMATOM Definition Flash running software version syngo CT2012BSOMATOM Definition Edge running software version syngo CT2012BSOMATOM Definition AS running software version syngo CT2012BSOMATOM Definition running software version syngo CT2012BSOMATOM Definition AS running software version syngo CT2013BARTG numbers: 164065 (SOMATOM system),160853 (syngo software)
  • Manufacturer

Manufacturer