SOMATOM Emotion 6 / 16 Slice Configuration and SOMATOM Spirit/Spirit Power Systems (Full-body computed tomography (CT) diagnostic x-ray system) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Siemens Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2014-RN-00457-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2014-04-17
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    The mechanical function of the rubber damper, which is situated within the mounting assembly of the rotation motor, may potentially degrade over time and may impact stability. in rare cases, the rubber damper may break and cause the loosening of the mounting plates in the gantry. this event is not posing a hazard by itself but could lead to a rupture of the drive belt and subsequently to a scan abort. additionally the broken belt may damage the system cover so that the contact protection against rotating parts inside the gantry can’t be granted for the operator and patient.
  • 조치
    Siemens is advising that there is no workaround for the user to minimise the potential risk of this issue. Siemens has developed a modification for the affected systems which is to replace the rotation motor mounting plate inside the gantry of the CT system by a new model. This new mounting plate includes significantly improved rubber dampers for the stability of the assembly. This action has been closed-out on 12/02/2016.