SOMATOM Emotion 6 / 16 Slice Configuration systems manufactured before February 2008 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Siemens Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2015-RN-00492-1
  • 사례 위험등급
    Class I
  • 사례 시작날짜
    2015-06-12
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    The three bolts used to fix the balance weight of the data measurement system (dms) may show signs of metal fatigue by producing an unusual noise whilst the gantry is in use and rotating. these bolts may potentially degrade over time and impact on stability of the balance weight. it should be noted that the rupture of a single bolt does not pose a hazard, however in the unlikely situation that all three bolts break, the dms balance weight is likely to drop into the gantry.This may cause serious patient injury and/or loss of scanned data and/or the machine rendered as out of service.No case reports have been received indicating patient injury.
  • 조치
    Service engineers will be replacing all the bolts holding the DMS balance weight with a new bolt design. In the interim, to minimise a potential hazard Siemens recommends for users to carefully monitor the noise generated during gantry rotation. Since bolts will deteriorate over time, the balance weight will slowly change its position, potentially leading to a new sound pattern. Should users become aware of a change to the typical noise of the scanner we recommend that you immediately stop operation and contact a Siemens service engineer. This action has been closed-out on 05/09/2016.