Sonopet Console (Ultrasonic Surgical Aspirator System) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Stryker Australia Pty Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2014-RN-00769-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2014-07-16
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    Stryker has noticed a trend of complaints regarding insufficient irrigation to the tip of the sonopet system. lack of irrigation can potentially lead to excessive heat at the tip. in the case where the irrigation pump fails, saline would not be provided to the tip. as a result the cooling function would be lost and could potentially cause the tip to heat up resulting in tissue burns to the patient.
  • 조치
    Stryker has updated the Instructions for Use (IFU) for Sonopet to include the following clarifications: · DO NOT use a rigid non-collapsible irrigation container. Failure to comply may cause the flow of irrigation to be drastically reduced and result in overheating of the handpiece tip and/or tip cover. Only use a collapsible irrigation container. · The irrigation tube must be properly routed through the irrigation pump with the pump cover fully closed and latched. Improper setup of the irrigation tubing may result in an inconsistent or lack of irrigation. · Provide adequate irrigation to the tip to avoid excessive heat that could result in tissue damage due to unintended contact. Additionally, all existing Sonopet consoles will have an instructional irrigation setup label applied to the side of the console. This action has been closed-out on 03/06/2016.

Device

  • 모델명 / 제조번호(시리얼번호)
    Sonopet Console (Ultrasonic Surgical Aspirator System)Catalogue Number: 5450852000All Serial NumbersARTG Number: 174364