Stereo Biopsy Device used with Mammomat Inspiration Mammography SystemMaterial number: 10140000 (Mammomat Inspiration) 의 리콜

Department of Health, Therapeutic Goods Administration에 따르면, 해당 리콜 는 Australia 에서 Siemens Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    RC-2014-RN-01040-1
  • 사례 위험등급
    Class II
  • 사례 시작날짜
    2014-09-25
  • 사례 국가
  • 사례 출처
    DHTGA
  • 사례 출처 URL
  • 비고 / 경고
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • 데이터 추가 비고
  • 원인
    The stereo biopsy devices for mammomat inspiration mammography systems might have integrated a safety switch which causes a failure of the functionality. the pin implemented in the safety switch may not put enough pressure on the safety circuit to prohibit movement. the needle positioning device may move even with the safety switch being set sideward. if this occurs with the needle being already inserted, it could lead to an unintentional patient injury.
  • 조치
    Customers are advised that in case the needle holder moves unintentionally to press the emergency stop. Siemens is recommending inspection of the safety switch before every biopsy until the problem is solved. Siemens is preparing a modification of the safety switch that will resolve this potential malfunction. The field modification will be available beginning of October. This action has been closed-out on 07/06/2016.

Device

  • 모델명 / 제조번호(시리얼번호)
    Stereo Biopsy Device used with Mammomat Inspiration Mammography SystemMaterial number: 10140000 (Mammomat Inspiration) Serial numbers: 1541, 2082ARTG number: 144223
  • Manufacturer

Manufacturer